The NeuroFortis is an automatic, motorized, washable drive. Compatible with all Alpha Omega systems.

Key Features

  • Compatible with Automated Washing Machine*
  • Meets Robert Koch Institute Recommendations for Cleaning and Sterilization of Medical Devices   
  • Preassembled with built-in recording and stimulation capability
  • Operated by remote control or manually
  • Drive with micrometer resolution feedback for precision movement
  • High amplifier input range allows non-stimulating channels to record throughout stimulation
  • Signal digitization and broadband filtering
  • 10 channels capable of recording unfiltered raw, high-frequency spike and low-frequency LFP signals**
  • Drives 5 microelectrodes simultaneously
  • New ergonomic knob design for user-friendly and accurate manual driving
  • A single cable connects the drive to all electrodes
  • A Single cable crossing the sterile field for all channels recording, and stimulation
  • High-quality material, durable design and production control for maximum durability in aggressive OR environment
  • XY stage and 5-hole Bengun allows trajectory adjustments without frame manipulation
  • Direct implantation of DBS electrode without altering the position of the drive or removing the cannulas
  • Lightweight, low impact on frameless or frame based procedure
  • Fully compatible with all stereotactic frames or frameless procedure

* Please follow the reprocessing instructions

**Available for NeuroOmega only. NeuroSmart is capable of recording 5 channels high-frequency spike and 5 channels low-frequency LFP signals no raw recording.

Range

40 mm

Resolution

100µm

Dynamic Input range

1mV

Input impedance

100MΩ/30pF

Frequency response (LFP Channels)

1Hz - 250Hz

Frequency response (Spike Channels)

400Hz-5.5KHz

Noise (1 Hz- 10KHz BW)

<8V P-P when inputs are shorted to GND

Input power

5V DC and 18V DC

Cleaning and Sterilization

Washable, Autoclavable

Case dimensions

110mm X 40mm X 50mm

Weight

190 Grams

The NeuroFortis can be used with various frame adaptors to fit a range of  Stereotactic frames,
The frame adaptor fits onto the Stereotactic frame and serves as the base on which to attach the NeuroFortis Drive Headstage.

  • Leksell frame
  • Leibinger ZD & RM Frames
  • CRW Frame
  • Nexframe
  • Starfix Frame
  • ROSA ONE Brain- robotic System
  • Renishaw – Neuromate Robotic System